Resumo
Aim: To assess the medium term outcomes of Acrysof® toric intraocular lens implantation in 54 patients (54 eyes). Methods: Uncorrected visual acuity (UCVA), best corrected visual acuity (BCVA), preoperative astigmatism, residual postoperative astigmatism, and global average and model-specific intraocular lens (IOL) rotation grade were analyzed. Results: At 2-months of follow-up,the mean UCVA was 0.83 (SD: 0.14) Snellen scale, with 73.9% of the patients >0.8, and 32.6% with 1.0. Mean BCVA achieved was 0.94 (SD: 0.10). Mean preoperatory astigmatism was -2.25 diopters (D) (SD: 0.78), and mean postoperative astigmatism was -0.32 D (SD: 0.56), with significant differences between both groups (p<0.001). Model-specific mean residual astigmatism was -0.1 D for T3, -0.27 D for T4 and -0.43 D for T5, without significant differences between the three models (p=0.483). Mean IOL-axis rotation grade was 3.87±3.25 degrees, with 91.6% of implanted lens within 10º of predicted axis. Discussion: T3, T4 and T5 Acrysof® Toric intraocular lenses can correct preoperative astigmatism with a high success rate in terms of UCVA, and residual postoperative astigmatism, with minimum IOL-rotation grade at 2 months follow-up period.
Referencias bibliográficas
- Horn, J.D. (2007). Status of toric intraocular lenses. Curr Opin Ophthalmol.. 18. 58-61
- Bauer, N.J., de Vries, N.E., Webers, C.A., Hendrikse, F., Nuijts, R.M. (2008). Astigmatism management in cataract surgery with the AcrySof toric intraocular lens. J Cataract Refract Surg.. 34. 1483
- Mendicute, J., Irigoyen, C., Aramberri, J., Ondarra, A., Montés-Micó, R. (2008). Foldable toric intraocular lens for astigmatism correction in cataract patients. J Cataract Refract Surg.. 34. 601
- Zuberbuhler, B., Signer, T., Gale, R., Haefliger, E. (2008). Rotational stability of the AcrySof SA60TT toric intraocular lenses: a cohort study. BMC Ophthalmol.. 8. 8
- Chang, D.F. (2008). Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses. J Cataract Refract Surg.. 34. 1842
- Ruiz-Mesa, R., Carrasco-Sánchez, D., Díaz-Álvarez, S.B., Ruiz-Mateos, M.A., Ferrer-Blasco, T., Montés-Micó, R. (2009). Refractive lens exchange with foldable toric intraocular lens. Am J Ophthalmol.. 147. 990