Radioterapia intraoperatoria en quirófano con haces de electronesdescripción técnica y dosimétrica del acelerador dedicado Sordina LIAC

  1. Sendón del Río, José Ramón
  2. Ayala Lázaro, Rafael
  3. Gómez Cores, Susana
  4. García Hernández, María Jesús
  5. Polo Cezón, Ramón
  6. Jiménez Rojas, Rocío
  7. López Bote, Miguel Ángel
Revista:
Revista de Física Médica

ISSN: 1576-6632

Any de publicació: 2015

Volum: 16

Número: 1

Pàgines: 43-50

Tipus: Article

Altres publicacions en: Revista de Física Médica

Resum

In this work we show our experience during the commissioning of a mobile electron-beam accelerator dedicated to intraoperative radiation therapy in the theatre room. The linac is a Sordina LIAC 12 MeV model with a hard-docking applicator system. We describe the linac, the measurement methods and the specific dosimetry. The dosimetric behavior is also discussed. Differences with other applicator systems can be explained from the particular head design of the linac.